Aanderaa 4531D Dissolved Oxygen Sensor
Sensor Maintenance
Maintenance must be conducted regularly to ensure the function and accuracy of the sensor.
Maintenance frequency |
Description |
Every Two Weeks |
Use a wet cloth to carefully clean the sensing foil. Visually inspect for buildup of organic matter - if excessive conduct monthly cleaning maintenance regime. |
Monthly |
The Optode housing can be cleaned using a brush and clean water, being careful to avoid the sensing foil. |
Sensor Calibration
Calibration should be conducted by an Aanderaa certified calibration laboratory. Please contact GroAqua Support to arrange calibration.
Calibration frequency |
Description |
Yearly |
In order to ensure the optimum accuracy yearly calibration is recommended. However as long as the sensor foil has not been changed and the sampling frequency is moderate the Aanderaa Optode will normally provide stable data for years. |
Fault Finding
Check K-Mote is on, if LED is not flashing, follow the diagnostics steps on the K-Mote sensor diagnostics page.
Fault |
Possible Issues |
Solution |
Erratic sensor readings |
Fouling buildup on sensor foil |
Visually inspect sensor foil and clean as per the above cleaning methods. |
No sensor readings |
Power supply issues, cable/connector may be disconnected or damaged |
Inspect sensor cable for damage, inspect connectors on K-Mote to ensure securely plugged in, inspect K-Mote power supply to ensure no damage has occurred. Ensure solar panel (if using) is orientated for optimal sun exposure and clean if necessary. Ensure K-Mote antenna is installed properly and the K-Mote is switched on. |
Further Advice
Contact |
Email Address |
Support Link |
Contact purpose |
Krucial |
support@krucial.com |
Krucial Support |
For issues with K-Devices and troubleshooting support |
GrowAqua |
support@groaqua.io |
GroAqua Support |
For issues relating directly to sensor and for the supply of spare parts/controlled reference fluid, or for further information on cleaning/calibrating the sensor. |